> Valeria,
>
> if you haven't loaded your samples yet on the gel, just add 5 or 10ul
> 2M DTT, it will work fine. If you already have run the gel, don't
> panic (too much), you just did not reduce your samples. They will run
> at different sizes maybe. You also might see some addional bands due
> to cross-linking and inter-/intramolecular -SS- bridges.
>
> Best,
>
> Wo
In addition, some antibodies don't recognize reduced proteins - they need a
"native", folded protein configuration.
--
Best regards